Specifications
37
Page 73
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Simulation compared to DCA Measurement
2.5G Data Rate through 20” trace
DCA Measurement of BERT
Simulation with Measured Data
Gain set to 1.90
9.1-9.49.4pS9.39pSJitter rms
37.1-38.537.1pS40.8pSJitter p-p
344-345.8344mV346mVEye Amp
172-176172pS157.9pSFall Time
169-178178pS157.9pSRise Time
-158.6-157.4-157.7mV-158.9mVLevel 0
186.8-188.6186mV187mVLevel 1
DCA
min/max
DCA typFrontPanel
[NOTE: FrontPanel data is from a single trace.
DCA measurements are averaged with 16
measurements.]
[NOTE: FrontPanel results are preliminary until
ADS 2005A final release.]
Page 74
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Simulation compared to DCA Measurement
5G Data Rate through 20” trace
DCA Measurement of BERT
Simulation with Measured Data
Gain set to 1.90
12.43/12.6512.5pS12.3pSJitter rms
53.28/54.0254.02pS54pSJitter p-p
293.4/293.8293.7mV292mVEye Amp
140.6/140.6140.6pS133pSFall Time
136.2/136.9136.2pS133pSRise Time
-136.6/-136-136mV-137mVLevel 0
156.4/157.5157.5mV155mVLevel 1
DCA
min/max
DCA typFrontPanel
[NOTE: FrontPanel data is from a single trace.
DCA measurements are averaged with 16
measurements.]
[NOTE: FrontPanel results are preliminary until
ADS 2005A final release.]