User manual
Operating Procedures 4
N9360A Multi UE Tester GSM User Manual 4-19
LossRfIn = Setting value at the LOSS/RF IN field on the [Configuration]
screen.
AttIn = Setting value at the
ATT In field on the [Configuration: Test
Sequence] screen.
There is a restriction that the sum of the LossRfIn and
ATT In must be
set more than or equal to 0.
Entering Loss in Configuration
Determine and enter the loss values caused by the antenna
coupler, RF cable, or shield case used to connect the mobile
phone to the Tester. If the Loss is set to
On, these path loss
values are applied to all through the test flow for the radio
system being currently tested.
Correction of input/output signal is described in Actual
Input/Output Level and Correction on page 17. Also, refer to Entering
Channel Attenuations on the Configuration: Test Sequence Screen on
page 20 for the correction of RF test in Automatic Test.
Step 1
Press the GSM softkey on the [Top Menu] screen to activate the
GSM system. Refer to Selection of System on page 13 for detail.
Step 2
Press the Config softkey on the [Initial] screen to display the
[Configuration] screen.