User manual
5-180 N9360A Multi UE Tester GSM User Manual
5 Screen Reference
BS Level
‡
Sets the RF output level of the Tester except for the Sensitivity test.
The allowable range is from -110.0 to -50.0 dBm in 0.1 dB steps.
This setting is for Automatic Test, Manual Test and Signal Generator.
BER BS
Level
‡
(GSM mode)
Sets the RF output level of the Tester for the Sensitivity test. The allowable range is from
-110.0 to -50.0 dBm in 0.1 dB steps.
This setting is for Automatic Test and Manual Test (GSM mode).
BLER BS Level
‡
(GPRS/EPGRS mode)
Sets the RF output level of the Tester for the Sensitivity test. The allowable range is from
-110.0 to -50.0 dBm in 0.1 dB steps.
This setting is for Manual Test (GPRS/EGPRS mode).
BER Frames
‡
(GSM mode)
Sets the number of test frames. The allowable range is from 1 to 13000.
The number of test bits, derived from the equation Test Frames multiplied by 78 bits, is
displayed in the parentheses.
This setting is for Automatic Test and Manual Test (GSM mode).
Fast BER
Frames
‡
(GSM mode)
Sets the number of test frames. The allowable range is from 1 to 13000.
The number of test bits, derived from the equation Test Frames multiplied by 456 bits, is
displayed in the parentheses.
This setting is for Manual Test (GSM mode).
BLER
Frames
‡
(GPRS/EPGRS mode)
Sets the number of test frames. The allowable range is from 1 to 13000.
The number of test bits, derived from the equation Test Frames multiplied by 78 bits, is
displayed in the parentheses.
This setting is for Manual Test (GPRS/EPGRS mode).
BER Frames
‡
(GPRS mode)
Sets the number of test frames for Test Mode B BER. The allowable range is from 1 to 13000.
The number of test bits, derived from the following equation is displayed in the parentheses.
Coding Scheme Equation
CS1
CS2
CS3
CS4
BER Frames x 160 bits
BER Frames x 240 bits
BER Frames x 288 bits
BER Frames x 400 bits
Loopback Delay
(GSM Mode)
Selects voice loop-back delay from Short, Mid and Long.
This setting is for Automatic Test and Manual Test.
Signaling Pattern
‡
Sets a signaling pattern. The allowable range is from 1 to 255.
Select 1 at present because settings over 2 will be extended in future.
Connection Wait
(GSM mode)
Sets the connection wait time for MS CALL.
The allowable range is from 0 to 9 sec in 1 sec step.
This setting is for Manual Test (GSM mode).
LocUpdate (Attach) Delay
‡
(GSM/GPRS/EGPRS
mode)
Sets the delay time after Location Update or Attach. The allowable range is from 0 to 9 sec
in 1 sec step.
This setting is for Manual and Automatic Test.
Table 5-78 [Configuration: Test Sequence] Screen Input Field
Input Field Description