Screen Reference 5
N9360A Multi UE Tester GSM User Manual 5-171
Reference Displays the current reference signal source.
The reference signal source cannot be changed on this screen. Change this on the [Configuration]
screen obtained from the [Top Menu] screen. Refer to Top Menu Screen on page 3.
• Internal: Uses the internal 10 MHz reference signal.
• External: Uses the external 10 MHz reference signal. The external 10MHz reference signal must be
supplied to the 10 MHz reference IN connector on the rear panel.
Serial Port Sets the serial (RS-232C) port communication condition to use external control. This function is
unavailable without E00/E01 option.
• Baud Rate: Sets communication speed to 9600, 19200, 38400, 57600 or 115200 bits per
second.
• Data Length: Sets data bit length to 7 or 8 bits.
• Stop Bits: Sets stop bit length to 1, 1.5 or 2 bits.
• Parity: Sets parity check mode to None, Odd or Even.
• Xcontrol: Sets flow control to None or Xon/Xoff.
Ethernet
*
• IP Addr: Sets IP Address. The allowable range is from 0.0.0.0 to 255.255.255.255.
• Mask: Sets Subnet mask. The allowable range is from 0.0.0.0 to 255.255.255.255.
Reboot the Tester after changing Ethernet setting.
GP-IB
*
Sets GP-IB. This function is unavailable without E00/E01 option.
• Address: Sets the address. The allowable range is from 1 to 15.
• EOI: Sets EOI control. Select On or Off.
• On: Enables EOI control. (Terminate character is invalid).
• Off: Disables EOI control. (Terminate character is valid).
Reboot the Tester after changing GP-IB setting.
Terminator Sets a terminator of output text data to CR, LF or CR+LF. This is the Terminator outputted from the
Tester. This setting is used for remote control with Serial, Ethernet and GP-IB.
(The Terminator of input text data is LF)
Date/ Time
*
Sets the current date and time in the following formats:
YYYY: Sets the year. The allowable range is from 1990 to 2037.
MM: Sets the month. The allowable range is from 01 to 12.
DD: Sets the day. The allowable range is from 01 to 31.
HH: Sets the hour. The allowable range is from 00 to 23.
MM: Sets the minute. The allowable range is from 00 to 59.
Table 5-74 [Configuration] Screen Input Field
Input Field Description