5-164 N9360A Multi UE Tester GSM User Manual
5 Screen Reference
The Cont mode setting provides
continuous measurement.
When the trigger mode is changed from
Cont to Sing, the measurement is
terminated.
Trigger: Starts single measurement.
More (1 of 2): Displays the softkey menu 2.
Return: Returns to the previous screen.
Softkey Menu 2
Print Screen: Prints a hardcopy of the screen or saves a graphic
file of the screen in a USB memory device
depending on the setting of the Printer input
field on the [Configuration] screen.
RF Output On/Off: Selects the RF power output from
On and
Off. RF On is displayed at the lower left
corner on screen when this is set to
On.
More (2 of 2): Returns to the softkey menu 1.
Screen Fields
The measurement results Table 5-71 as is shown.
Set the input fields of the following parameters to execute a new
test.
Table 5-71 [TX Analyzer] Spectrum Monitor (span +-100kHz) Screen Field
Screen Field Description
Peak The absolute peak power and frequency are shown.
Marker
*
The readout of frequency and amplitude at the marker position is shown as follows when the
marker is activated by the Knob Cur/Marker softkey setting:
——— MHz —— dBm: The absolute frequency and amplitude at the marker position are
shown.
(—— kHz —— dB): The relative frequency and amplitude the marker position are shown.
M1: –67.7kHz
M2: 0.0kHz
M3: +67.7kHz
The peak amplitudes around M1 (–67.7 kHz), M2 (0.0 kHz), and M3 (+67.7 kHz) relative to the
RFCH frequency offsets are shown.
* The resolution of the marker on the screen is not equivalent to the measurement resolution.