5-162 N9360A Multi UE Tester GSM User Manual
5 Screen Reference
•The Cont mode setting provides
continuous measurement.
When the trigger mode is changed from
Cont to Sing, the measurement is
terminated.
• Trigger: Starts single measurement.
• More (1 of 2): Displays the softkey menu 2.
• Return: Returns to the previous screen.
Softkey Menu 2
• Print Screen: Prints a hardcopy of the screen or saves a graphic
file of the screen in a USB memory device
depending on the setting of the Printer input
field on the [Configuration] screen.
• RF Output On/Off: Selects the RF power output from
On and
Off. RF On is displayed at the lower left
corner on screen when this is set to
On.
• More (2 of 2): Returns to the softkey menu 1.
Screen Field
The measurement results as Table 5-70 is shown.
Figure 5-72 is the [Stand-by] Spectrum Monitor screen for span
+-100 kHz.
Table 5-70 [TX Analyzer] Spectrum Monitor (span 0 - 400kHz) Screen Field
Screen Field Description
Peak The absolute peak power and frequency are shown.
Marker
*
The readout of frequency and amplitude at the marker position is shown as follows when the
marker is activated by the Knob Cur/Marker softkey setting:
• ——— MHz —— dBm: The absolute frequency and amplitude at the marker position are
shown.
• (—— kHz —— dB): The relative frequency and amplitude the marker position are shown.
M1: +200kHz
M2: +250kHz
The amplitude at +200 kHz and +250 kHz offsets relative to the RFCH frequency are shown.
* The resolution of the marker on the screen is not equivalent to the measurement resolution.