User manual

Screen Reference 5
N9360A Multi UE Tester GSM User Manual 5-151
When the trigger mode is changed from
Cont to Sing, the measurement is
terminated.
Trigger: Starts single measurement.
More (1 of 2): Displays the softkey menu 2.
Return: Returns to the previous screen.
Softkey Menu 2
Print Screen: Prints a hardcopy of the screen or saves a graphic
file of the screen in a USB memory device
depending on the setting of the Printer input
field on the [Configuration] screen.
RF Output On/Off: Selects the RF power output from
On and
Off. RF On is displayed at the lower left
corner on screen when this is set to
On.
More (2 of 2): Returns to the softkey menu 1.
Screen Field
Set the input fields of the following parameters to execute a new
test.
Table 5-64 [TX Analyzer] Phase Error/Frequency Error Screen Input Field
Input Field Description
Procedure Displays a list of parameter file names stored in the Tester.
If the Tester is using a parameter file, the file name is displayed in this field. If the Tester contains
parameter files, a list of those filename is displayed in this field for selection.
Radio System Selects a radio system from GSM850, GSM900, DCS1800 and PCS1900.
RFCH
*
Sets the channel number.
The allowable range is as follows:
GSM850: 128 to 251
GSM900: 0 to 124 and 955 to 1023
DCS1800: 512 to 885
PCS1900: 512 to 810
Signal Selects CW or Burst of a MS transmission signal.
CW: Selects this when a continuous wave is transmitted from the mobile phone.
Burst: Selects this when a bursted signal is transmitted from the mobile phone.