User manual
Screen Reference 5
N9360A Multi UE Tester GSM User Manual 5-149
The measurement results are shown in the screen field.
Phase Error/ Frequency Error
The RMS and Peak Phase Error and Frequency Error are shown
along with the marker readout if the marker is activated.
If the measured data exceeds the limits on the graph, one or
more white triangle markers indicate the failed position at the
top of the graph frame.
Any data indicating a failure is highlighted.
NOTE
If the measurement is not triggered, check the following items:
•
Power Control Level (PWR CNTL) is properly set for the mobile phone.
• Midamble is properly set. If you don't know which Training Sequence Code
(TSC) is used, set Midamble to None to use the amplitude trigger mode.
A Pass/Fail judgment is not given while transmission power is off.
Table 5-63 [TX Analyzer] Peak TX Power/Burst Timing/Power Ramp Screen Field
Screen Field Description
Peak TX Power Measured Peak TX Power, the HI and the LO Limit are shown. The HI and the LO Limit are set on
the [Configuration: Test Condition] screen.
Burst Timing Measured Burst Timing, the HI and the LO Limit are shown. The HI and the LO Limit are set on
the [Configuration: Test Condition] screen.
Power Ramp Measured Power Ramp, the HI and the LO Limit are shown. The limits are specified by the radio
system selected (template).