5-126 N9360A Multi UE Tester GSM User Manual
5 Screen Reference
•To terminate continuous measurement on
the [Measuring] screen, change the trigger
mode from
Cont to Sing.
•This softkey is effective only for changing
the trigger mode on the [Stand-by] screen.
• Trigger: Starts single measurement. This softkey is inactivated
on the [Stand-by] screen.
• More (1 of 2): Displays the softkey menu 2.
• Return: Returns to the previous screen.
Softkey Menu 2
• Print Screen: Prints a hardcopy of the screen or saves a graphic
file of the screen in a USB memory device
depending on the setting of the Printer input
field on the [Configuration] screen.
• More (2 of 2): Returns to the softkey menu 1.
Screen Field
Set the input fields of the following parameters to execute a new
test.
Table 5-52 [Manual Test] EGPRS Mode EVM Screen Input Field
Input Field Description
Procedure Displays a list of parameter file names stored in the Tester.
Select a file to recall for a test.
If the Tester is using a parameter file, the file name is displayed in this field. If the Tester contains
parameter files, a list of those filename is displayed in this field for selection.
This field cannot be changed on a [Measuring] screen.
Radio System Selects a radio system from GSM850, GSM900, DCS1800 and PCS1900.
This field cannot be changed on the [Measuring] screen.
BCCH
*
Sets the broadcast control channel number.
The allowable range is as follows:
GSM850: 128 to 251
GSM900: 0 to 124 and 955 to 1023
DCS1800: 512 to 885
PCS1900: 512 to 810
This field cannot be changed on the [Measuring] screen.