User manual
5-124 N9360A Multi UE Tester GSM User Manual
5 Screen Reference
EVM
The RMS, Peak, 95 %, Frequency Error, and Origin Offset are
shown along with the marker readout if the marker is activated.
If the measured data exceeds the limits on the graph, one or
more white triangle markers indicate the failed position at the
top of the graph frame.
Any data indicating a failure is highlighted.
Table 5-51 [Manual Test] EGPRS Mode Phase Error/Frequency Error Screen Field
Screen Field Description
Phase RMS Measured Phase Error (RMS) and the HI Limit are shown. The HI Limit is set on the
[Configuration: Test Condition] screen.
Phase Peak Measured Phase Error (Peak), the HI and the LO Limit are shown. A set of the HI and the LO
Limit is set on the [Configuration: Test Condition] screen.
Marker Measured Phase Error at the marker position, the HI and the LO Limit are shown. Measured data
is shown when the marker is placed on the measurement trace. The HI Limit is set on the
[Configuration: Test Condition] screen.
Freq. Err Measured Frequency Error, the HI and the LO Limit are shown. The HI and the LO Limit are set on
the [Configuration: Test Condition] screen.
NOTE
Phase Error/Frequency Error function is valid when the Coding Scheme is
set to MCS1 to MCS4 on the [Standby] Screen.
NOTE
Install the Option G03 to display EVM screen (EGPRS mode). Set the
Coding Scheme to MCS5 to MCS9 on the [Standby] Screen.