User manual

5-120 N9360A Multi UE Tester GSM User Manual
5 Screen Reference
The measurement results are shown in the screen field.
BCCH
*
Sets the broadcast control channel number.
The allowable range is as follows:
GSM850: 128 to 251
GSM900: 0 to 124 and 955 to 1023
DCS1800: 512 to 885
PCS1900: 512 to 810
This field cannot be changed on the [Measuring] screen.
PDTCH
*
Sets a packet data traffic channel number.
The allowable range is as follows:
GSM850: 128 to 251
GSM900: 0 to 124 and 955 to 1023
DCS1800: 512 to 885
PCS1900: 512 to 810
PWR CNTL Sets one of the factors defined for Power Control Levels according to the Radio System. The allowable
range is as follows:
GSM850, GSM900: 2 to 19 (+39dBm to +5dBm)
DCS1800: 29 to 15 (+36dBm to 0dBm)
PCS1900: 30 to 15 (+33dBm to 0dBm)
Set the MS Power Class on the [Configuration: Test Condition] screen for the power class of the mobile
phone. If the mobile phone does not support the power control function, measurement result is not
displayed.
BS Level
*
Sets the RF output level of the Tester, ranging from -110.0 to -50.0 dBm in 0.1 dB step. The
default value is set on the [Configuration: Test Condition] screen.
Averaging
*
Sets the averaging number from 2 to 99 or Off. When this function is not required, select Off.
* The changing magnification softkey is available. Refer to Storing Numeric Values on page 8 and Changing Magnification Softkey on
page 13.
Table 5-48 [Manual Test] EGPRS Mode Peak TX Power/ Burst Timing/Power Ramp Screen Input Field
Input Field Description
NOTE
A Pass/Fail judgment is not given while transmission power is off.
Table 5-49 [Manual Test] EGPRS Mode Peak TX Power/Burst Timing/Power Ramp Screen Field
Measurement Item Description
Peak TX Power Measured Peak TX Power, the HI and the LO Limit are shown. The HI and the LO Limit are set on
the [Configuration: Test Condition] screen.