User manual

Screen Reference 5
N9360A Multi UE Tester GSM User Manual 5-47
The Sing mode setting provides single
measurement in RF test.
The
Cont mode setting provides
continuous measurement in RF test.
Trigger: Starts measurement in the connection status. This
softkey is inactivated in the stand-by status but
activated in the connection status.
More (1 of 2): Displays the softkey menu 2.
Return: Returns to the [Initial] screen.
Softkey Menu 2
Print Screen: Prints a hardcopy of the screen or saves a graphic
file of the screen in a USB memory device
depending on the setting of the Printer input
field on the [Configuration] screen.
Spectrum Monitor: Displays the Spectrum Monitor screen.
SMS: Displays the SMS screen.
More (2 of 2): Returns to the softkey menu 1.
Screen Field
Set the input fields in the screen field according to the following
description of Table 5-13 to execute a new test.
Table 5-13 [Manual Test: Stand-by] Screen GSM Mode Input Field
Input Field Description
Measuring
mode
Sets whether or not to execute each group of test items. Select On or Off.
Peak TX Power, Burst Timing, Power Ramp
Phase Error (RMS), Phase Error (Peak), Frequency Error
Sens/BER II, Sens/FER (Sens: sensitivity)
Procedure Displays a list of parameter file names stored in the Tester.
Select a file to recall for a test.
If the Tester is using a parameter file, the file name is displayed in this field. If the Tester contains
parameter files, a list of those filename is displayed in this field for selection.
Radio System Selects a radio system from GSM850, GSM900, DCS1800 or PCS1900.