User manual

5-30 N9360A Multi UE Tester GSM User Manual
5 Screen Reference
Softkey Menu 3
Simple: Displays a Pass/Fail result of whole measurement in
the middle of the measurement result field in zoom.
Detail: Displays a Pass/Fail result at each item cell in the
measurement result field.
Value: Displays a measurement value at each item cell in the
measurement result field.
Cancel: Cancels changing the display mode and returns to the
softkey menu 1.
Screen Field
The information in the screen field is displayed. Refer to
Table 5-3 [Automatic Test: Stand-by] Screen Input Field on
page 19 and Table 5-6 [Automatic Test: Measuring] Screen
Mobile Phone Information on page 25. The input field in the
screen field can be changed for a new test.
Explanation
Detailed measurement results up to the aborted operation can
be obtained for further analysis from the Detail screen and the
Value screen. Place the circular cursor at one of the Pass/Fail
result or result value cells and press the CURSOR CONTROL
knob to display the Graph/Value screen containing detailed
measurement results.
The status of the Tester is reset to stand-by.
The condition of transmission from the mobile phone is not
ensured; therefore, the mobile phone power-on process needs to
be repeated for a new test.
Sequence 2 Screen
When Sequence 2 is set on the [Configuration: Test Sequence]
screen in the Configuration mode, Sequence 2 start after
completion of BS Release in Sequence 1.
The state of the sequences is shown on the upper left screen.
The sequence in processing (Sequence 2 now) is underlined.
Test result of sequences is shown as follows:
P: Shown on blue background when the test is passed.