User manual
Screen Reference 5
N9360A Multi UE Tester GSM User Manual 5-19
Softkey Menu 2
• Print Screen: Prints a hardcopy of the screen or saves a graphic
file of the screen in a USB memory device
depending on the setting of the Printer input
field on the [Configuration] screen.
• More (2 of 2): Returns to the softkey menu 1.
Softkey Menu 3
• Simple: Displays a Pass or Fail result of whole measurement
in the middle of the measurement result field in
zoom.
• Detail: Displays a Pass or Fail result at each item cell in the
measurement result field.
• Value: Displays a measurement value at each item cell in the
measurement result field.
• Cancel: Cancels selecting the display mode and returns to the
softkey menu 1.
Screen Fields
The input fields of parameters except procedure are protected if
the Panel Key is set to
Lock on the [Configuration] screen.
Set the input fields in the screen field according to the following
description of Table 5-3 to execute a new test.
Table 5-3 [Automatic Test: Stand-by] Screen Input Field
Input Field Description
Procedure Displays a list of parameter file names stored in the Tester.
Select a file to recall for a test.
If the Tester is using a parameter file, the file name is displayed in this field. If the Tester contains
parameter files, a list of those filename is displayed in this field for selection.
BCCH
*
Sets the broadcast control channel number.
The allowable range is as follows:
GSM850: 128 to 251
GSM900: 0 to 124 and 955 to 1023
DCS1800: 512 to 885
PCS1900: 512 to 810