User manual
5-8 N9360A Multi UE Tester GSM User Manual
5 Screen Reference
Also, the following information is shown in the screen field:
• MAC Address
• Serial No.
• Option: Displays options in numerical orders. This field
remains blank when no option is installed in the Tester.
GP-IB
*
Sets GP-IB. This function is unavailable without E00/E01 option.
• Address: Sets the address. The allowable range is from 1 to 15.
• EOI: Sets EOI control. Select On or Off.
• On: Enables EOI control. (Terminate character is invalid).
• Off: Disables EOI control. (Terminate character is valid).
Reboot the Tester after changing GP-IB setting.
Terminator Sets a terminator of output text data to CR, LF or CR+LF. This setting is used for remote control with
Serial, Ethernet and GP-IB.
(The Terminator of input text data is always LF)
Date/ Time
*
Sets the current date and time in the following formats:
• YYYY: Sets the year. The allowable range is from 1990 to 2037.
• MM: Sets the month. The allowable range is from 01 to 12.
• DD: Sets the day. The allowable range is from 01 to 31.
• HH: Sets the hour. The allowable range is from 00 to 23.
• MM: Sets the minute. The allowable range is from 00 to 59.
Printer Sets the output direction of print screen to USB Memory or EPSON PM-G800.
• USB Memory: Saves a graphic file of screen in a USB memory device. The image format is PNG.
• EPSON PM-G800: Prints a hardcopy of screen in the specified printer.
Beeper Sets beep.
Selects On or Off.
• On: Beeps for each step of operation.
• Off: Beeps only for noticing some errors and warnings.
AutoBoot Selects the system booted specified time after the [Top Menu] screen is displayed.
• Application: Sets the system from the followings
• None: Not automatically boots
• GSM: Boots the GSM option
• WCDMA: Boots the W-CDMA option
• Timer: Sets the timeout from 10 to 60 seconds, or -. The timer is stopped if any softkey is pressed.
• —: Invalid when Application is set to None.
• 10 to 60: Specifies timeout when Application is set to GSM or WCDMA.
* The changing magnification softkey is available. Refer to Changing Magnification Softkey on page 13.
Table 5-1 [Configuration] Screen Input Field
Input Field Description