N9360A Multi UE Tester GSM User Manual 5-7
When the Option E01 is not installed in the Tester, the Serial Port
GP-IB field is displayed in gray (inactivated) as Figure 5-6.
Screen Field
The following input fields need to be set to use the Tester
according to the following description of Table 5-1.
Figure 5-5 Part of [Configuration] screen (without Option E00)
Figure 5-6 Part of [Configuration] screen (without Option E01)
Table 5-1 [Configuration] Screen Input Field
Input Field Description
Reference Selects the reference signal from Internal and External.
• Internal: Use the internal 10 MHz reference signal.
• External: Use the external 10 MHz reference signal. The 10 MHz reference signal must be supplied
to the 10MHz Reference IN connector on the rear panel.
Serial Port Sets the serial (RS-232C) port communication condition to use external control. This function is
unavailable without E00/E01 option.
• Baud Rate: Sets communication speed to 9600, 19200, 38400, 57600 or 115200 bits per
second.
• Data Length: Sets data length to 7 or 8 bits.
• Stop Bits: Sets stop bit length to 1, 1.5 or 2 bits.
• Parity: Sets parity check mode to None, Odd or Even.
• Xcontrol: Sets flow control to None or Xon/Xoff.
Ethernet
*
• IP Addr.: Sets IP Address. The allowable range is from 0.0.0.0 to 255.255.255.255.
• Mask: Sets Subnet mask. The allowable range is from 0.0.0.0 to 255.255.255.255.
Reboot the Tester after changing Ethernet setting.