User manual
7-8 N9360A Multi UE Tester CDMA2000 User Manual
7 Performance and Specifications
Maximum TX Power Measurement
Maximum TX Power Measurement specifications are shown on
Table 7-14.
Minimum TX Power Measurement
Minimum TX Power Measurement specifications are shown on
Table 7-15.
Table 7-14 Maximum TX Power Measurement
Item Specification Unit Remarks
Measurement Range –60 to +36 dBm
Resolution 0.01 dB
Measurement Accuracy –53 to +36 dBm
±1.5 (25 °C ±5 °C)
±1.5 (0 °C ±50 °C)
dB
Downlink Output Level
: <–50 dBm
–60 to –53.01 dBm
±1.5 (25 °C ±5 °C)
±2.0 (0 °C ±50 °C)
dB
Table 7-15 Minimum TX Power Measurement
Item Specification Unit Remarks
Measurement Range –60 to +36 dBm
Resolution 0.01 dB
Measurement Accuracy –53 to +36 dBm
±1.0 (25 °C ±5 °C)
±1.0 (0 °C ±50 °C)
dB
Downlink Output Level
: <–50 dBm
–60 to –53.01 dBm
±1.5 (25 °C ±5 °C)
±2.0 (0 °C ±50 °C)
dB