Table 7-1 Power Measurement
Item Specification Unit Remarks
Measurement range –60 to +36 dBm
Resolution 0.01 dB
Measurement accuracy –53 to +36 dBm Downlink Output Level
±1.0 (25 °C ± 5 °C) dB : <
50 dBm
±1.5 (0 °C ± 50 °C)
–60 to –53.01 dBm
±1.5 (25 °C ± 5 °C) dB
±2.0 (0 °C ± 50 °C
Table 7-2 Frequency Error Measurement
Item Specification Unit Remarks
Input Level –20 to +36 dBm
Measurement Range ±10.0 kHz
Resolution 0.1 Hz
Measurement Accuracy ±(30.0 + frequency accuracy of the reference signal) Hz