Error codes displayed on error screens
Table 6-2 and Table 6-3 list error codes for [Error] screens. If
the following guidelines do not help you to remove the errors,
contact the Agilent Sales Department or an authorized
distributor.
Table 6-2 Alarm Notification Error Code
Error Code Error Content Action
0000 Abnormal internal temperature Turn off power and disconnect the AC plug and contact our sales
department or authorized distributor.
0001 The cooling fan stopped.
0002 Synthesizer unlock (Tx1) The internal reference signal generation circuit may be out of order.
Turn off power and disconnect the AC code and contact our sales
department or authorized distributor. If you use the external reference
signal, check if it is connected correctly.
0003 Synthesizer unlock (Tx2)
0004 Synthesizer unlock (Tx3)
0005 Synthesizer unlock (Rx1)
0006 Synthesizer unlock (Rx2)
0007 ALC alarm (Tx1) A hardware error occurred. Reboot the Tester.
0008 ALC alarm (Tx2)
1000 Unit reboot is detected. A firmware error occurred. Reboot the Tester.
1001 Unit fatal error
1002 Unit initialization failure
1005 No indispensable file
1010 Receiving NAK from RF-CPU
2000 No MS configuration
information
2002 No scenario
2003 Scenario read error