Description of Screens 5
N9360A Multi UE Tester CDMA2000 User Manual 5-171
* A changing magnification softkey is available. Refer to Storing Numeric Values on page 4-10 and Changing Magnification Softkey on
page 4-13.
Configuration: File Management
When the File Management softkey is pressed on the
[Configuration] screen, the screen as Figure 5-75 is displayed.
All parameter files are listed in the table.
You can save or recall a parameter file to/from the following two
devices:
USB memory
HDD in the Tester
Select the device you want to use by pressing the HDD/USB
softkey.
Sensitivity/PER Test 3 (95%) This item is displayed only when the 3GPP2 Config is set to 3 or 4 (1xEV-DO
Rev. A). Sets high measurement limit of the Sensitivity/PER Test 3 when the
Confidence is set to 95%. The allowable setting range is from 0.0 to 5.0% (0.5%
steps).
Sensitivity/PER Test 3 (Off) This item is displayed only when the 3GPP2 Config is set to 3 or 4 (1xEV-DO
Rev. A). Sets high measurement limit of the Sensitivity/PER Test 3 when the
Confidence is set to Off. The allowable setting range is from 0.00 to 100.00%
(0.01% steps)
Table 5-52[Configuration: Test Condition (Limit)] Measurement Item (1xEV-DO)
Measurement Item Description
NOTE
If you use a USB memory device, insert a proper USB memory device into
the USB Connector on the front panel of the Tester.