User manual

5-170 N9360A Multi UE Tester CDMA2000 User Manual
5 Description of Screens
Before entering the test limits, make sure that the CDMA Mode
and the frequency band are selected correctly for the test limits
you are entering.
Table 5-52[Configuration: Test Condition (Limit)] Measurement Item (1xEV-DO)
Measurement Item Description
CDMA Mode Sets the CDMA Mode to MC-1x or 1xEV-DO.
Band Sets the Band to either of:
B0:Cel US
B1:PCS US
B3:Cel JP
B4:PCS KR
Access Probe Power Sets high and low measurement limits of the Access Probe Power. The
allowable range is from –99.9 to +99.9 dBm in 0.1 dB steps.
Inner Loop Power(ILP) Sets high and low measurement limits of the Inner Loop Power. The allowable
range is from –99.9 to +99.9 dB in 0.01 dB steps.
Max TX Power Sets high and low measurement limits of the Maximum Transmission Power.
The allowable range is from –99.9 to +99.9 dBm in 0.01 dB steps.
Min TX Power Sets high and low measurement limits of the Minimum Transmission Power.
The allowable range is from –99.9 to +99.9 dBm in 0.01 dB steps.
Frequency Error Sets high and low measurement limits of the Frequency Error. The allowable
range is from –10000.0 to +10000.0 Hz in 0.1 Hz steps.
Multi-code Rho Sets high and low measurement limits of the waveform quality (ρ). The
allowable range is from 0.800 to 1.000 (0.001 steps).
Origin Offset Sets high and low measurement limits of the Origin Offset. The allowable range
is from –99.9 to 0.0 dBc in 0.1 dB steps.
Time Offset Sets high and low measurement limits of the Time Offset. The allowable range
is from –10.00 to +10.00 sec in 0.01 sec steps.
Sensitivity/PER Test 1 (95%) Sets high measurement limit of the Sensitivity/PER Test 1 when the
Confidence is set to 95 %. The allowable setting range is from 0.0 to 5.0% (0.5%
steps).
Sensitivity/PER Test 1 (Off) Sets high measurement limit of the Sensitivity/PER Test 1 when the
Confidence is set to Off. The allowable setting range is from 0.00 to 100.00%
(0.01% steps).
Sensitivity/PER Test 2 (95%) Sets high measurement limit of the Sensitivity/PER Test 2 when the
Confidence is set to 95 %. The allowable setting range is from 0.0 to 5.0% (0.5%
steps).
Sensitivity/PER Test 2 (Off) Sets high measurement limit of the Sensitivity/PER Test 2 when the
Confidence is set to Off. The allowable setting range is from 0.00 to 100.00%
(0.01% steps).