5-168 N9360A Multi UE Tester CDMA2000 User Manual
5 Description of Screens
* A changing magnification softkey is available. Refer to Storing Numeric Values on page 4-10 and Changing Magnification Softkey on
page 4-13.
Table 5-51[Configuration: Test Condition (Limit)] Measurement Item (MC-1x)
Measurement Item Description
CDMA Mode Sets the CDMA Mode to MC-1x or 1xEV-DO.
Band Sets the Band to either of:
B0: Cel US
B1: PCS US
B3: Cel JP
B4: PCS KR
B6: IMT-2K
Access Probe Power Sets high and low measurement limits of the Access Probe Power. The
allowable range is from –99.9 to +99.9 dBm in 0.1 dB steps.
Inner Loop Power (ILP) Sets high and low measurement limits of the Inner Loop Power. The allowable
range is from –99.9 to +99.9 dB in 0.01 dB steps.
Max TX Power Sets high and low measurement limits of the Maximum Transmission Power.
The allowable range is from –99.9 to +99.9 dBm in 0.01 dB steps.
Min TX Power Sets high and low measurement limits of the Minimum Transmission Power.
The allowable range is from –99.9 to +99.9 dBm in 0.01 dB steps.
Frequency Error Sets high and low measurement limits of the Frequency Error. The allowable
range is from –10000.0 to +10000.0 Hz in 0.1 Hz steps.
Rho Sets high and low measurement limits of the waveform quality (ρ). The
allowable range is from 0.800 to 1.000 (0.001 steps).
Multi-code Rho Sets high and low measurement limits of the waveform quality (ρ). The
allowable range is from 0.800 to 1.000 (0.001 steps).
Origin Offset Sets high and low measurement limits of the Origin Offset. The allowable range
is from –99.9 to 0.0 dBc in 0.1 dB steps.
Time Offset Sets high and low measurement limits of the Time Offset. The allowable range
is from –10.00 to +10.00 sec in 0.01 sec steps.
Sensitivity/ FER (95%) Sets high measurement limit of the sensitivity/FER when the Confidence is set
to 95%. The allowable range is from 0.0 to 5.0% (0.5% steps).
Sensitivity/ FER (Off) Sets high measurement limit of the sensitivity/FER when the Confidence is set
to Off. The allowable range is from 0.00 to 100.00% (0.01% steps).
Sensitivity/TDSO FER (95%) Sets high measurement limit of the Sensitivity/TDSO FER when the Confidence
is set to 95 %. The allowable range is from 0.0 to 5.0% (0.5% steps).
Sensitivity/TDSO FER (Off) Sets high measurement limit of the Sensitivity/TDSO FER when the Confidence
is set to Off. The allowable range is from 0.00 to 100.00% (0.01% steps).