User manual

5-164 N9360A Multi UE Tester CDMA2000 User Manual
5 Description of Screens
Access Probe
*
Set parameters for the access probe.
Setting Open Loop Adj (Open Loop Adjust)
0 to -255 (1 dB step)
Setting Probe Init Adj (Probe Initial Adjust)
-16 to +15 (1 dB step)
Setting Probe Num Step (Probe Num Step)
1 to 15
Setting Power Step (Power Step)
0 to 15 (1 step) × 0.5 dB
If you turn on Access Probe Pwr Meas, Power Step is disabled and 0 is displayed.
When you turn off Access Probe Pwr Meas, Power Step is enabled. Values will be
restored (before they were set to ON) and displayed.
PER
*
Set parameters for FER measurement. Set items and the allowable range are as follows:
Max Packet
25 to 10000000 (10
7
) for a step
Confidence
Set either Off or 95%.
Access Probe Pwr Meas Set measurement of Access Probe Power.
On or Off
Averaging Specifies the average number of the measurement of the TX Power, Frequency Error,
Multi-code Rho, Origin Offset, Time Offset, Max TX Power, Min TX Power and Code Domain
Power measurement.
Sets the averaging number from 2 to 99, or Off.
Max TX Power BS Level
*
Set BS level to measure Max Power in a manual test.
The allowable range is between -115.0 and -18.0 dBm in 0.1 dB step.
Min TX Power BS Level
*
Set BS level to measure Min Power in a manual test.
The allowable range is between -115.0 and -18.0 dBm in 0.1 dB step.
Access Network
*
Set Access Network.
The allowable range by set item is as follows:
Country Code
0 to 4095 (1 step)
Sector ID
0 to FFFF(1 step Hex)
Subnet Mask
0 to FF(1 step Hex)
Access Terminal
*
UATI Color Code
0 to FF(1 step Hex)
UATI024
0 to FFFFFF(1 step Hex)
Table 5-49 [Configuration: Test Condition] Screen Input Field (1xEV-DO) (continued)
Input Field Description