Description of Screens 5
N9360A Multi UE Tester CDMA2000 User Manual 5-161
FER
*
Set parameters for FER measurement. Set items and the allowable range are as follows:
Max Frames
25 - 10000000 (107) for a step
Confidence
Set either Off or 95%.
Connection Wait
*
Set connection suspense time for MS Call in a manual test.
0: Immediate response
1 to 120: Waiting a response in seconds and specifying timeout Automatic response when
the specified time elapsed
When the time is set to 1-120 seconds, pressing the Answer softkey before an automatic
response changes to the manual response.
Averaging Specifies the average number of the measurement of the TX Power, Frequency Error,
Waveform Quality (Rho or multi-code Rho), Origin Offset, Time Offset, Max TX Power, Min TX
Power and Code Domain Power measurement.
Sets the averaging number from 2 to 99, or Off.
SCH Rate Set the SCH rate. This item is displayed only when Radio Conf is set to 3, 4 or 5. The
allowable range is as follows:
Radio Config: For 3
9.6, 19.2, 38.4, 76.8, 153.6kbps
Radio Config: For 4
9.6, 19.2, 38.4, 76.8, 153.6kbps
Radio Config: For 5
14.4, 28.8, 57.6, 115.2, 230.4.6kbps
SCH Encoder Set the SCH encode type. This item is displayed only when Radio Conf is set to 3, 4 or 5. It is
compulsorily set to CONV for SCH Data Rate=9.6k/14.4k.
CONV
TURB
When SCH Rate is set to 9.6 kbps or 14.4 kbps, the TURB cannot be selected.
SCH Level Sets code level of the forward supplemental Channel. This item is displayed only when the
Radio Config is set to F3R3, F4R3, or F5R4 with the Service Option of 32. The allowable
range is as follows:
–20.00 to –2.50 dB (in 0.25 dB steps)
Radio Conf Selects Radio Configuration. The allowable setting range is as follows:
F1R1 (Forward 1, Reverse 1)
F2R2 (Forward 2, Reverse 2)
F3R3 (Forward 3, Reverse 3)
F4R3 (Forward 4, Reverse 3)
F5R4 (Forward 5, Reverse 4)
When the F1R1 or F2R2 is selected, the waveform quality is displayed as Rho. When the
others are set, Multi-code Rho is displayed.
Loopback Delay Selects sound loop-back delay from Short, Mid and Long.
This setting is for Automatic Test and Manual Test.
Table 5-48 [Configuration: Test Condition] Screen Input Field (MC-1x) (continued)
Input Field Description