5-156 N9360A Multi UE Tester CDMA2000 User Manual
5 Description of Screens
BS Level (Band) Sets output signal level from the Tester for each band.
Band:
Selects the band to set the BS level.
0: Band 0 (Cel US)
1: Band 1 (PCS US)
3: Band 3 (Cel JP)
4: Band 4 (PCS KR)
6: Band 6 (IMT-2K)
BS Level:
Sets the BS level for the protocol test and the Access Probe Power measurement.
–115.0 to –18.0 dBm (0.1 dB steps)
Measurement BS Level:
Sets the BS level for the measurement of Inner Loop Power, Frequency Error, Multi-code Rho and Time
Offset.
–115.0 to –18.0 dBm (0.1 dB steps)
Max TX Power BS Level:
Sets the BS level for the Max TX Power measurement.
–115.0 to –18.0 dBm (0.1 dB steps)
Min TX Power BS Level:
Sets the BS level for the Min TX Power measurement.
–115.0 to –18.0 dBm (0.1 dB steps)
PER BS Level 1:
Sets the BS level for the PER1 measurement.
–115.0 to –18.0 dBm (0.1 dB steps)
PER BS Level 2:
Sets the BS level for the PER2 measurement.
–115.0 to –18.0 dBm (0.1 dB steps)
PER BS Level 3: (Available when the "3GPP2 Config" is set to 3 or 4 (1xEV-DO Rev. A))
Sets the BS level for the PER3 measurement.
–115.0 to –18.0 dBm (0.1 dB steps)
RF Output Sets the RF output for Automatic Test. Select Auto or On.
Auto: Outputs RF signal only during measurement.
On: Outputs RF signal while an [Auto Test] screen is displayed.
UATI
Assignment
Timeout
Set Timeout time for UATI Assignment Timeout. The allowable range is between 5 to 300 seconds
(1 step).
Table 5-46 Input items on the [Configuration: Test Sequence] (continued)Screen (1xEV-DO)
Input Item Description