Description of Screens 5
N9360A Multi UE Tester CDMA2000 User Manual 5-151
BS Level (Band) Sets output signal level from the Tester for each band.
Band:
Selects the band to set the BS level.
0: Band 0 (Cel US)
1: Band 1 (PCS US)
3: Band 3 (Cel JP)
4: Band 4 (PCS KR)
6: Band 6 (IMT-2K)
BS Level:
Sets the BS level for the protocol test and the Access Probe Power measurement.
–115.0 to –18.0 dBm (0.1 dB steps)
Measurement BS Level:
Sets the BS level for the measurement of Inner Loop Power, Frequency Error, Rho or Multi-code Rho
and Time Offset.
–115.0 to –18.0 dBm (0.1 dB steps)
Max TX Power BS Level:
Sets the BS level for the Max TX Power measurement.
–115.0 to –18.0 dBm (0.1 dB steps)
Min TX Power BS Level:
Sets the BS level for the Min TX Power measurement.
–115.0 to –18.0 dBm (0.1 dB steps)
FER BS Level:
Sets the BS level for the FER measurement.
–115.0 to –18.0 dBm (0.1 dB steps)
RF Output Sets the RF output for Automatic Test. Select Auto or On.
Auto: Outputs RF signal only during measurement.
On: Outputs RF signal while an [Auto Test] screen is displayed.
Location Update
Timeout
Set Timeout time for Location Update. The allowable range is between 5 and 300 seconds (1 step).
Table 5-43 Input items on the [Configuration: Test Sequence] (continued)Screen (MC-1x)