User manual

5-118 N9360A Multi UE Tester CDMA2000 User Manual
5 Description of Screens
RFCH
*
Specifies the band and channel number. This parameter can be changed in the stand-by status.
The allowable range of channel number for each band is as follows:
B0 (Cel US): 1024 to 1323CH, 991 to 1023CH, 1 to 799CH
B1 (PCS US): 0 to 1199CH
B3 (Cel JP): 1041 to 1199CH, 1201 to 1600CH, 801 to 1039CH, 1 to 799CH
B4 (PCS KR): 0 to 599CH
B6 (IMT-2K): 0 to 1199CH
Refer to Appendix B General information on cdma2000 system for details.
Amplitude Sets the RF output level of the tester between -115.0 and -18.0 dB at 0.1dB step.
Modulation Specifies the Modulation from the Modulation Data setting and the Modulation Mode setting.
Modulation Data setting:
MC-1x Mode
Pilot Only
Pilot+Sync+Pch+OCNS+Fch
Pilot+Sync+Pch+OCNS+Fch+Sch
1xEV-DO Mode
Cont Pilot
Pilot+MAC+Cntrl
Pilot+MAC+Cntrl+Tch
Other
AM Modulation 1
AM Modulation 2
Modulation Mode setting (Select the modulation mode from the followings when the modulation data is
set to Pilot+MAC+Cntrl+Tch on the 1xEV-DO mode):
QPSK
8PSK
16QAM
The selectable modulation systems depend on the mode you have selected. For the allowable
combinations of Mode and Modulation system, refer to Appendix D Parameters of physical channels.
Averaging Specifies the average number of the measurement of the TX Power, Frequency Error, Multi-code Rho,
Origin Offset, Time Offset and Code Domain Power measurement.
Sets the averaging number from 2 to 99, or Off.
* The changing magnification softkey is available. Refer to Storing Numeric Values on page 4-10 and Changing Magnification Softkey on
page 4-13.
Table 5-32 Entries on the [TX Analyzer: Stand-by] Screen (continued)
Input Item Description