User manual
5-110 N9360A Multi UE Tester CDMA2000 User Manual
5 Description of Screens
Message Field
The following message is displayed in the message field on the
[Measuring] screen:
Press the desired softkey.
Table 5-30 Mobile Phone Information on the [Manual Test: Measuring] Screen (1xEV-DO)
Item Information Description
ESN Displays the Electronic Serial Number sent from the mobile phone, with maximum 8 digits.
Pilot Strength Displays the pilot channel measured level at the mobile phone between –13.5 and 0.0 dB when the
mobile phone receives the Measurement Report message.
Access Probe Power Measures and displays the access probe TX power of the mobile phone.
The display range is –60.0 to 36.0 dBm
• This is not measured and not displayed when the Access Probe Meas field on the
[Configuration: Test Condition] screen is set to OFF.
3GPP-SYS Displays the value of the 3GPP System that is set on the [Configuration: Test Sequence] and the
[Configuration: Test Condition] screens.
SIG-PTN Displays the value of the Signaling Pattern that is set on the [Configuration: Test Condition]
screen.
3GPP2-CONF Displays the value of the 3GPP2-Config that is set on the [Configuration: Test Condition] screen.