Description of Screens 5
N9360A Multi UE Tester CDMA2000 User Manual 5-107
Softkey Menu 2
Print Screen: Prints a hardcopy of the screen or saves a
graphic file of the screen in a USB memory device
depending on the setting of the Printer input field on the
[Configuration] screen.
Call Drop Timer On/Off: Set Call Drop Timer to On or Off.
More (2 of 2): Returns to the Softkey Menu 1.
Screen Field
Set the input fields in the screen field according to the following
description of Table 5-28 to execute a new test.
Table 5-28 Entries on the [Manual Test: Measuring] Screen for Code Domain Power Measurement (MC-1x)
Input Item Description
Measuring
mode
Sets On (measure) or Off (not measured) for each test item group or measurement item.
TX Power
Pilot, PRI, DRC, ACK, Data, DSC
C
, Max Inactive Channel
The DSC is displayed only when the 3GPP2 Config is set to 3 or 4 (1xEV-DO Rev. A).
If Code Domain Power test item is set to On, the TX Power measurement is also set to On. If the test
item is set to Off, the TX Power measurement is also set to Off.
RFCH
*
Specifies the band and channel number. This parameter can be changed when the stand-by or
Connection status.
The allowable range of channel number for each band is as follows:
B0 (Cel US): 1024 to 1323CH, 991 to 1023CH, 1 to 799CH
B1 (PCS US): 0 to 1199CH
B3 (Cel JP): 1041 to 1199CH, 1201 to 1600CH, 801 to 1039CH, 1 to 799CH
B4 (PCS KR): 0 to 599CH
B6 (IMT-2K): 0 to 1199CH
Setting RFCH in the Connection status starts handoff. Handoff is also available between bands.
Refer to Appendix B General information on cdma2000 system for details.
BS Level
*
Sets the RF output level of the Tester, ranging from -115.0 to -18.0 dBm in 0.1 dB step. The
default value is set on the [Configuration: Test Condition] screen.
Per BS Level
*
Sets the RF output level of the Tester for the packet error rate measurement. The allowable range is from
–115.0 to –18.0 dBm in 0.1 dB steps. The default value is set on [Configuration: Test Condition] screen.
PWR CNTL
*
Specifies the power control bit to control TX Power of the mobile phone. The allowable setting is as
follows:
HOLD
CNT UP (Continuous up)
CNT DWN (Continuous down)
Max Packets
*
Specifies the number of packets used for PER measurement. The allowable range is from 25 to
10000000 (or 10