Description of Screens 5
N9360A Multi UE Tester CDMA2000 User Manual 5-99
The information of Table 5-25 sent from a mobile phone is
displayed on the screen.
BS Level
*
Sets the RF output level of the Tester, ranging from -115.0 to -18.0 dBm in 0.1 dB step. The
default value is set on the [Configuration: Test Condition] screen.
PER FT Rate Sets the Forward Data Rate (FT Rate) for the packet error rate measurement. The selectable values are:
• 307.2 kbps
• 2457.6 kbps
• 3072.0 kbps (This can be set only when Release A is set in 1xEV-DO mode if 3GPP2 Config is 3 or
4.)
PWR CNTL Specifies the power control bit to control TX Power of the mobile phone. The allowable setting is as
follows:
• HOLD
• CNT UP (Continuous up)
• CNT DWN (Continuous down)
Max Packets
*
Specifies the number of packets to measure the PER measurement. The allowable range is from 25 to
10000000 (or 10
7
).
Confidence Sets the confidence level of the PER measurement. Selects 95% or Off.
Averaging Specifies the average number of the measurement of the TX Power, Frequency Error, Multi-code Rho,
Origin Offset, Time Offset, Max TX Power, Min TX Power and Code Domain Power measurement.
Sets the averaging number from 2 to 99, or Off.
* The changing magnification softkey is available. Refer to Storing Numeric Values on page 4-10 and Changing Magnification Softkey on
page 4-13.
Table 5-24 Entries on the [Manual Test: Stand-by] Screen (1xEV-DO)
Input Field Description
Table 5-25 Information from a Mobile Phone displayed on the [Manual Test: Stand-by] Screen (1xEV-DO)
Item Information Description
ESN Displays the Electronic Serial Number sent from the mobile phone, with Max. 8 digits. This
number cannot be displayed when the state is Stand-by.
Pilot Strength Displays the pilot channel measured level at the mobile phone between –13.5 and 0.0 dB when
the mobile phone receives the Measurement Report message.
Radio System The CDMA2000 message is displayed.
Access Probe Power When an access probe is received, its power is measured and displayed before connection. The
display range is from –60.0 to 36.0 dBm.
• This is not measured and not displayed when the Access Probe Meas field on the
[Configuration: Test Condition] screen is set to OFF.
3GPP-SYS Displays the value of the 3GPP System that is set on the [Configuration: Test Sequence] and the
[Configuration: Test Condition] screens.