User manual

5-94 N9360A Multi UE Tester CDMA2000 User Manual
5 Description of Screens
Manual Test (1xEV-DO Mode)
This section explains the standard manual test procedure of a
mobile phone in the 1xEV-DO Manual Test mode.
Overview
Features of Manual Test
All test items are executed while the mobile phone and the
Tester are connected by call control protocol.
Measurement in the RF test step in the test flow is executed
in a real-time.
The following systems can be selected directly from the
Manual Test screen.
Required options are in parentheses.
GSM (G00), GPRS (G00), EGPRS (G00, G03), W-CDMA (W00,
W06), HSDPA (W00, W06), cdma2000 MC-1x (C00, C01).
Setting for Manual Test in the Configuration Mode
Loss:
The path loss values for each band can be set on the
[Configuration: Test Condition (Loss)] screen to correct the
RF attenuation between the mobile phone and the Tester.
Refer to Actual Input/Output Level and Correction on page 4-19 for
correction of input/output signal.
Limits of measurement items:
Each measurement item can be measured and analyzed with
test limits set on the [Configuration: Test Condition (Limit)]
screen.
NOTE
Install the C02 option to use the test function in the 1xEV-D0 mode.