User manual

Description of Screens 5
N9360A Multi UE Tester CDMA2000 User Manual 5-89
Softkey Menu 1
Screen>>: Displays the softkey menu 2 containing the
softkeys to select the display mode.
Abort: Aborts the measurement and returns to the
[Stand-by] screen.
Softkey Menu 2
Simple: Displays a Pass/Fail result of whole measurement
in the middle of the measurement result field in zoom.
Detail: Displays a Pass/Fail result at each item cell in the
measurement result field.
Value: Displays a measurement value at each item cell in
the measurement result field.
Cancel: Cancels changing the display mode and returns to
the Softkey Menu 1.
Result Screen
When all tests in preset sequences are completed, measurement
results are displayed on the [Stand-by] Result screen.
The results are displayed on the [Automatic Test: Stand-by]
Result Simplified Screen (1xEV-DO) (Figure 5-45), the
[Automatic Test: Stand-by] Result Detailed Screen (1xEV-DO)
(Figure 5-46) or the [Automatic Test: Stand-by] Result Value
Screen (1xEV-DO) (Figure 5-47) depending on the previous
setting of the Simple, the Detail or the Value softkey.
Measurement results are shown as follows:
Pass or measurement value: Displayed on blue background
when the measurement is passed.
Fail or measurement value: Displayed on red background
when the measurement is failed.
Measurement result is shown depending on the display mode.
On the result simplified display screen, an enlarged
Pass is
shown in the center of the measurement result field when all
measurement items are passed. An enlarged
Fail is shown
when even one measurement item is failed.