User manual
Description of Screens 5
N9360A Multi UE Tester CDMA2000 User Manual 5-77
• Sequence:
Up to two test sequences can be set on the [Configuration:
Test Sequence] screen.
RF Test in Automatic Test
Table 5-17 shows the measurement items of the RF test. For the
judgment criteria of measurement items, refer to Configuration:
Test Condition Screen.
Table 5-17 explains the measurement items for Automatic Test.
The result (pass or fail) of each measurement item is set on the
[Configuration: Test Condition] screen.
Table 5-17 Measurement Items for Automatic Test (1xEV-DO)
Measurement Item Description
Access Probe Power Measures and displays the transmission power of the
access probe from the mobile phone..
ILP (Down) Measures the transmission power difference of the
mobile phone when the Tester controls the
transmission power of the mobile phone down to 10
dB.
ILP (Up) Measures the transmission power difference of the
mobile phone when the Tester Control the transmission
power of the mobile phone up to 10 dB.
Max TX Power Measures the maximum transmission power of the
mobile phone.
Min TX Power Measures the minimum transmission power of the
mobile phone.
Frequency Error Measures a transmission frequency error of the mobile
phone.
Multi-code Rho Measures the modulation accuracy (ρ) of transmission
waveforms of the mobile phone.
Time Offset Measures a transmission timing error of the mobile
phone.
Sensitivity/PER1,
Sensitivity/PER2
Sensitivity/PER3
Measures the packet error rate.
Sensitivity/PER3 is displayed in the RF Test item only
when the 3GPP2-CONF is set to 3 or 4 (ReleaseA).