User manual
Description of Screens 5
N9360A Multi UE Tester CDMA2000 User Manual 5-67
Softkey Menu Field
Softkey Menu 1
• Trigger Sing/Cont: Measures the Code Domain Power. Select
Sing (single measurement) or Cont (continuous
measurement) as the trigger mode. This item is
unavailable on this screen.
• Selecting
Cont as the trigger mode will immediately
start the continuous measurement.
• Changing the trigger mode from
Cont to Sing will end
the measurement.
• Trigger: Measures the Code Domain Power. Selecting
Sing
as the trigger mode will start the measurement after
pressed.
• More (1 of 2): Displays the Softkey Menu 2.
• Return: Returns to the [Connection] screen.
Softkey Menu 2
• Print Screen: Prints the hard-copy of the screen or stores
the screen image in the USB memory, according to the
Printer field on the [Configuration] screen.
• More (1 of 2): Returns to the Softkey Menu 1.
Screen Field
Set the input items required to execute tests, referring to the
description shown in Table 5-15.
Table 5-15 Input Items on [Manual Test: Measuring] Code Domain Power Measurement Screen (MC-1x)
Input Field Description
Measuring
mode
Sets On (measured) or Off (not measured) for each test item group or measurement item.
• TX Power
• Pilot, Traffic (Fundamental), Max Inactive Channel
TX Power Measurement is unconditionally conducted at execution of Code Power Measurement, which
indicates that Pilot/Traffic (Fundamental) Measurement is set to On.
• If Pilot/Traffic (Fundamental) Measurement is set to On, TX Power Measurement is also set to On. If
it is set to Off, TX Power Measurement is also set to Off.