5-64 N9360A Multi UE Tester CDMA2000 User Manual
5 Description of Screens
The information from the mobile phone as Table 5-14 is
displayed on the left side of the screen.
RFCH
*
Specifies the band and channel number. This parameter can be changed when the stand-by or
Connection status.
The allowable range of channel number for each band is as follows:
• B0:Cel US 1024 to 1323CH, 991 to 1023CH, 1 to 799CH
• B1:PCS US 0 to 1199CH
• B3:Cel JP 1041 to 1199CH, 1201 to 1600CH, 801 to 1039CH, 1 to 799CH
• B4:PCS KR 0 to 599CH
• B6:IMT-2K 0 to 1199CH
• Setting RFCH in the Connection status starts handoff. Handoff is also available between bands. Refer
to Appendix B General information on cdma2000 system for details.
RX Power Sets On (measure) or Off (not measure) for RX Power measurement.
When this parameter set to on during the Stand-by status and starts the MS call or the BS Call, measure
the RX power once and the Tester display the result.
BS Level
*
Sets the RF output level of the Tester, ranging from -110.0 to -18.0 dBm in 0.1 dB step. The
default value is set on the [Configuration: Test Condition] screen.
Averaging Specifies the average number of the measurement of the TX Power, Frequency Error, Waveform Quality
(Rho or multi-code Rho), Origin Offset, Time Offset, Max TX Power, Min TX Power and Code Domain
Power.
Sets the averaging number from 2 to 99, or Off.
PWR CNTL
*
Specifies the power control bit to control TX Power of the mobile phone. The allowable setting is as
follows:
HOLD: Sets to HOLD.
UP: Increases the TX Power of the mobile phone to the specified value. (Not available in the Stand-by
status.)
DOWN: Decreases the TX Power of the mobile phone to the specified value. (Not available in the
Stand-by status.)
CNT UP: Sets to Continuous Up.
CNT DWN: Sets to Continuous Down.
This softkey is not available when the Service Option is set to 1, 3, 17 or 32768.
Max Frames
*
Specify the number of frames to measure Frame Error Ratio.
The allowable range is as follows: 25 to 10000000.
Confidence Selects confidence level of the Frame Error Ratio Measurement. Select 95% or Off.
* The changing magnification softkey is available. Refer to Storing Numeric Values on page 4-10 and Changing Magnification Softkey on
page 4-13.
Table 5-13 Input Items on [Manual Test] Connection Screen (MC-1x)
Input Field Description