5-30 N9360A Multi UE Tester CDMA2000 User Manual
5 Description of Screens
Automatic RF Test
Table 5-4 shows the measurement items of the RF test. For the
judgment criteria of measurement items, refer to Table 5-49.
Table 5-4 Measurement Items for Automatic Test (MC-1x)
Measurement Item Description
Access Probe Power Measures and displays the access probe power when receiving an access probe before
connection.
ILP (Down) Measures the transmission power difference of the mobile phone when the Tester controls
the transmission power of the mobile phone down to 10 dB.
ILP (Up) Measures the transmission power difference of the mobile phone when the Tester controls
the transmission power of the mobile phone up to 10 dB.
Max TX Power Measures the maximum transmission power of the mobile phone.
Min TX Power Measures the minimum transmission power of the mobile phone.
Frequency Error Measures the transmission frequency error of the mobile phone.
Rho/Multi-code Rho Measures the transmission wave modulation accuracy (ρ) for the mobile phone.
Time Offset Measures the transmission timing error of the mobile phone.
Sensitivity/FER Measures the frame error rate.
NOTE
A changing magnification softkey to input numerical values will be
enabled. Refer to Storing Numeric Values on page 4-10 and Changing
Magnification Softkey on page 4-13.
Default settings for the Tester at the time when the power is turned on
represent the settings that were set last time the power was turned off.