User manual
Description of Screens 5
N9360A Multi UE Tester CDMA2000 User Manual 5-29
Automatic Test (MC-1x Mode)
Overview
Feature of Automatic Test
• Pass or Fail test results in the test flow and measurement
items are automatically obtained.
• This function executes RF test changing frequencies on up to
six channels automatically. Combining Sequence1 and
Sequence2 enables measurement up to 12 channels in one
test.
Setting for Automatic Test in Configuration Mode
• Parameters:
The parameters used in Automatic Test are set on the
[Configuration: Test Sequence] and [Configuration: Test
Condition] screens.
• Loss:
The path loss values for each band can be set on the
[Configuration] screen to correct the RF attenuation
between the mobile phone and the Tester. Refer to Actual
Input/Output Level and Correction on page 4-19 for correction of
input/output signal.
• Test limits:
The test limits for each test can be defined on the
[Configuration: Test Condition (Limit)] screen. The test
result (pass or fail) for each test is automatically judged by
using the predefined test limits.
• ATT In/Out:
The path loss values for each traffic channel can be set on the
[Configuration: Test Sequence] screen to correct the RF tests
results and RF output level at each traffic channel in each
band.
• Sequence:
Up to two test sequences can be set on the [Configuration:
Test Sequence] screen.