User manual

5-8 N9360A Multi UE Tester CDMA2000 User Manual
5 Description of Screens
Screen Field
Set the input items required to use the Tester, referring to the
description shown in Table 5-1.
Table 5-1 Input Items on the [Configuration] Screen
Input Field Description
Reference Selects the reference signal from Internal and External.
Internal: Use the internal 10 MHz reference signal.
External: Use the external 10 MHz reference signal. The 10 MHz reference signal must be supplied
to the 10MHz Reference IN connector on the rear panel.
Serial Port Sets the serial (RS-232C) port communication condition to use external control. This function is
unavailable without E01/E02 option.
Baud Rate: Sets communication speed to 9600, 19200, 38400, 57600 or 115200 bits per
second.
Data Length: Sets data length to 7 or 8 bits.
Stop Bits: Sets stop bit length to 1, 1.5 or 2 bits.
Parity: Sets parity check mode to None, Odd or Even.
Xcontrol: Sets flow control to None or Xon/Xoff.
Ethernet
*
Set various settings for external control via an Ethernet port.
IP Addr: Specifies the IP address.
Mask: Specifies the net mask.
Determine whether the Tester should be controlled via the router or gateway installed outside.
None: Controls the Tester without using the router and gateway.
ACT: Controls the Tester using the router or gateway.
Addr: Determine whether the Tester should be controlled via the router or gateway installed
outside. This item is available when Gateway is set to ACT.
Reboot the Tester after changing Ethernet setting.
GP-IB
*
Sets the communication condition of the GP-IB port to use remote control. The option E02 is required
to use this function.
Addr: Sets the address. The allowable range is from 1 to 15.
EOI: Sets EOI control. Select On or Off.
On: Enables EOI control. (Terminator setting is ignored).
Off: Disables EOI control. (Terminator setting is valid).
Reboot the Tester after changing GP-IB setting.
Terminator Sets a terminator of output text data to CR, LF or CR+LF. This setting is used for external control
with Serial, Ethernet and GP-IB.
(The Terminator of input text data is always LF)
Printer Set the destination to which the print data is sent when it is printed (when the Print Screen softkey is
pressed) to USB Memory or EPSON PM-G800.
USB Memory: Saves a graphic file of screen in a USB memory device. The image format is PNG
(Portable Network Graphics).
EPSON PM-G800: Prints a hardcopy of the screen in the specified printer.