User manual

Operating Procedures 4
N9360A Multi UE Tester W-CDMA User Manual 4-43
13 At the "Connection (AMR)" step, the Tester returns voice
back to the mobile phone. Check the voice quality.
14 Also, CPICH RSCP can be measured at the "Connection
(AMR)" step. Press the Measurement CPICH RSCP softkey to
make the mobile phone report the value of CPICH RSCP,
and then the value is displayed.
15 Finish the call. Press an On Hook button on the mobile
phone to start MS Release, or press the Release softkey to
start BS Release.
16 Press the BS Call softkey to start BS Call (AMR). PRACH
Power is measured. Also, CPICH RSCP is reported from
the mobile phone when the "CPICH RSCP" field is set to
"On".
17 Respond to the call from the Tester. Press the Off Hook
button on the mobile phone.
18 At the "Connection (AMR)" step, the Tester returns voice
back to the mobile phone. Check voice quality.
19 Also, CPICH RSCP can be measured at the "Connection
(AMR)" step.
Press the Measurement CPICH RSCP softkey to make the
mobile phone report the value of CPICH RSCP, and then
the value is displayed.
20 Finish the call. Press the Release softkey to start BS
Release or press an On Hook button on the mobile phone
to start MS Release.
21 Press the BS Call softkey to start BS Call (RMC). RFCH
Power is measured. Also, CPICH RSCP is reported from
the mobile phone when the "CPICH RSCP" field is set to
"On".
22 The mobile phone automatically responds to the call from
the tester.
23 At the "Connection (RMC)" step, RF Test is executed. Start
the measurement with the following softkey operation.
Trigger Sing/Cont : When the trigger mode is Sing, single
measurement starts when the Trigger softkey is pressed.
To change the trigger mode to continuous mode (Cont),
press the Trigger Sing/Cont softkey. When the trigger
mode is Cont, continuous measurement starts as soon
as the "Connection (RMC)" step starts.
Trigger: Starts single measurement.