User manual

7-2 N9360A Multi UE Tester W-CDMA User Manual
7 Performance and Specifications
Measurement Performance
Power Measurement
Table 7-1 shows the specification of the power measurement.
Table 7-1 Power Measurement
Item Specification Unit Remarks
Range -60 to +36 dBm
Measurement resolution 0.01 dB
Display resolution Open Loop 0.1
dB
Inner Loop 0.01
MAX TX Power 0.01
PRACH Power 0.1
Accuracy
(MAX TX Power, TX Power)
0 to +36 dBm Downlink Output Level:
–50dBm
± 0.7 (25ºC ± 5ºC)
± 1.0 (0ºC to 50ºC)
dB
–53 to –0.01 dBm
± 1.0 (25ºC ± 5ºC)
± 1.5 (0ºC to 50ºC)
dB
–60 to –53.01 dBm
± 1.5 (25ºC ± 5ºC)
± 2.0 (0ºC to 50ºC)
dB
Accuracy
(Inner Loop)
–20 to 36 dBm Downlink Output Level:
–50dBm
1dB Up/Down:±0.2
10dB Up/Down: ±0.5
dB