Troubleshooting 6
N9360A Multi UE Tester W-CDMA User Manual 6-5
Error Codes for Error Screens
Table 6-2 and Table 6-3 list all the error codes for the error
screens. If the following guidelines do not help you to remove
the errors, contact the
Agilent Sales Department or an authorized
distributor
.
Table 6-2 Alarm Notification Error Code
Error Code Error Content Action
0000 Abnormal internal temperature Turn off power and disconnect the AC plug.
0001 The Cooling Fan stopped.
0002 Synthesizer unlock (Tx1) A failure of the internal reference oscillator may have occurred. Turn
off power and disconnect the AC cord. If you use the external
reference signal, check if it is connected correctly.
0003 Synthesizer unlock (Tx2)
0004 Synthesizer unlock (Tx3)
0005 Synthesizer unlock (Rx1)
0006 Synthesizer unlock (Rx2)
0007 ALC alarm (Tx1) A hardware error occurred. Reboot the Tester.
0008 ALC alarm (Tx2)
1000 Unit reboot is detected. A firmware error occurred. Reboot the Tester.
1001 Unit fatal error
1002 Unit initialization failure
1003 The combination of UI and SC
*
is not correct.
1005 No indispensable file.
1010 Receiving NAK from RF-CPU
2000 No MS configuration
information
2002 No scenario
2003 Scenario read error
3001 Scenario execution failure
because of a hardware error
* SC: an application module in the Tester