5-166 N9360A Multi UE Tester W-CDMA User Manual
5 Screen Reference
Softkey Menu Field
Softkey Menu
Print Screen: Prints a hardcopy of the screen or saves a graphic
file of the screen in a USB memory device depending on the
setting of the "Printer" input field on the [Configuration] screen.
Return: Returns to [Configuration] screen.
Screen Fields
Set the input fields in the screen field according to the following
description of Table 5-51 to execute a new test.
† The changing magnification softkey is available. Refer to “Storing Numeric Values" on
page 4-10 and “Changing Magnification Softkey" on page 4-13.
Table 5-51 [Configuration: Test Condition (Loss)] Screen Input Field
Input Field Description
Loss † Sets the path losses caused by the coupler or cable used to connect the RF signals from/to the
mobile phone with each radio systems, respectively.
Selects "On" or "Off".
On: Adds a set path loss value of RF In/Out. Select "On" to activate the change of ATT
setting on the [Configuration: Test Sequence] screen.
Off: Not add the path loss value.
Depending on transmitter testing and receiver testing, set loss values in the "RF In" and "RF
Out" fields as follows:
RF In: Sets the traffic channel loss values for transmitter testing. The allowable range is
from 0.0 to 99.9 dB in 0.1 dB steps.
RF Out: Sets the traffic channel loss for receiver testing. The allowable range is from 0.0 to
99.9 dB in 0.1 dB steps.
The sum of Loss setting value and ATT setting value has some limitation. Refer to“Ac t ua l
Input/Output Level and Correction" on page 4-19.