† The changing magnification softkey is available. Refer to “Storing Numeric Values" on
page 4-10 and “Changing Magnification Softkey" on page 4-13.
AUTH Key Sets the type of Authentication Key to "ORG", "STD", or "USER".
• ORG: Sets it to Test USIM provided by Agilent.
• STD: Sets it to the specified value of 3GPP.
• USER: Sets the type of USIM to others.
The default value for each type is as follows.
• ORG: 00112233445566778899AABBCCDDEEFF
• STD: 4147494C454E5420544543484E4F0000
• USER: 4147494C454E5420544543484E4F0000
For Test USIM (p/n:1150-8007) used "STD"
For Test USIM (p/n:E5515-10007/10008) used "USER"
Authentication Key can be set by remote control only when Authentication Key is set to "USER".
Refer to the "N9360A Multi UE Tester W-CDMA Pragramming Manual" for details.
LO/HI Limit Sets Pass/Fail acceptance values of RF test. Refer to "Setting Measurement Limits on
Measurement Items" on next page.
Table 5-49 [Configuration: Test Condition] Screen Input Field