page 4-10 and “Changing Magnification Softkey" on page 4-13.
Setting Measurement Limits on Measurement Items
The "LO Limit" and "HI Limit" values for the measurement items
in Table 5-48 need to be set to execute Pass/Fail tests with
Automatic Test, Manual Test and TX Analyzer. Limits can be
specified in all measurement items of test set on the
[Configuration: Test Sequence] screen (even if some test items
are skipped).
The changing magnification softkey is available. Refer to
“Storing Numeric Values" on page 4-10 and “Changing
Magnification Softkey" on page 4-13.
AUTH Key Sets the type of Authentication Key to "ORG", "STD", or "USER".
ORG: Sets it to Test USIM provided by Agilent.
STD: Sets it to the specified value of 3GPP.
USER: Sets the type of USIM to others.
A default value for each type is as follows.
ORG: 00112233445566778899AABBCCDDEEFF
STD: 4147494C454E5420544543484E4F0000
USER: 4147494C454E5420544543484E4F0000
For Test USIM (p/n:1150-8007) used "STD"
For Test USIM (p/n:E5515-10007/10008) used "USER"
Authentication Key can be set by remote control only when the Key is set to "USER". Refer to
N9360-90703 Agilent N9360A Multi UE Tester W-CDMA Programming Manual for details.
LO/HI Limit Sets Pass/Fail acceptance values of RF test.
Refer to “Setting Measurement Limits on Measurement Items" on page 5-160 in the following
section.
Table 5-47 [Configuration: Test Condition] Screen Input Field
Input Field Description
Table 5-48 [Configuration: Test Condition] Screen Test Limits
Measurement Items Description
Open Loop TX Power Sets a high and a low test limit.
The allowable range is from -99.9 to +99.9dBm in 0.1dB steps.
ILP (Down Min) Sets a high and a low test limit.
The allowable range is from -99.99 to +99.99dB in 0.01 dB steps.
ILP (Down Max) Sets a high and a low test limit.
The allowable range is from -99.99 to +99.99dB in 0.01 dB steps.