BS Level (Band) † • Band: Sets the Band of BS Level which will be set.
• BS Level: Sets a RF output level at signaling (not at measurement). The allowable range is
from -115.0 to -18.0 dBm in 0.1 dB steps.
• Measurement BS Level: Sets a RF output level at MAX TX Power, Inner Loop Power and EVM
measurement. The allowable range is from -115.0 to -18.0 dBm in 0.1 dB steps.
• Openloop BS Level: Sets a RF output level at Open Loop TX Power measurement. The
allowable range is from -115.0 to -18.0 dBm in 0.1 dB steps.
• FreqError BS Level: Sets a RF output level at Frequency Error measurement. The allowable
range is from -115.0 to -18.0 dBm in 0.1 dB steps.
• BER BS Level: Sets a RF output level at BER measurement. The allowable range is from -115.0
to -18.0 dBm in 0.1 dB steps.
Test Items Sets execution or skip for each signaling test step. Refer to “A u tomati c Tes t " on page 5-30.
RF Output Sets the RF output control for Automatic Test. Select "Auto" or "On".
• Auto: Outputs RF signal only during measurement.
• On: Outputs RF signal while and [Automatic Test] screen is displayed.
BER Frames † Sets the number of test frames for BER measurement. The allowable range is from 1 to 4100 in 1
step. (1 Frame = 1 TTI)
Also, a bit number according to BER Frames is displayed between parentheses. The bit number
is calculated by an equation "BER Frames multiplied by 244". While BER Frames is entered, The
bit number is displayed in conjunction with the value of BER Frames.
Wait Before Paging † Sets the wait time before paging transmission during Open Loop TX Power measurement. The
allowable range is from 0.0 (no wait) to 99.9 seconds in 0.1second step.
3GPP System Selects 3GPP version and the protocol. Contact the Agilent Sales Department about
corresponding mobile phones.
The parameter is 1, 2, 3, or 4.
Table 5-46 [Configuration: Test Sequence] Screen Input Field