User manual

5-112 N9360A Multi UE Tester W-CDMA User Manual
5 Screen Reference
PWR CNTL Sets control for power control level of the mobile phone to "UP", "DOWN", "CNT UP", or "CNT
DWN" and also a control value.
HOLD: Sets HOLD.
Up: Increases the power control value in the set control value.
Down: Decreases the power control value in the set control value.
CNT UP: Increases the control value continuously.
CNT DWN: decreases the power control value continuously.
The setting is done with the following operations.
When this field is changed from "HOLD" to "UP" or "DOWN", this setting is automatically
returned to "HOLD" after completion of this setting.
When this field is set to "CNT UP" or "CNT DWN", the setting is remained after release the
call.
When this field is changed from "CNT UP" to "CNT DWN", this setting is remained as "CNT
UP" after completion of this setting. (vice versa)
Set TPC Algorithm at "TPC Algorithm" field on the [Configuration: Test Condition] screen.
The allowable range for the control value is from 1 dB to 99 dB in 1 dB step.
Ec/Ior Sets the code power of the DPCH, HS-SCCH, HS-PDSCH by fixed combination.
The available combination is defined by the combination of the FRC Type, Modulation Type, UE
Category.
The HS-PDSCH code power stated above is the total amount of HS-PDSCH code power of each
Channlisation Code.
Refer to Ta ble 5-2 9.
FRC Type Sets the FRC Type and each modulation type that can be set.
Allowable range is as follows:
FRC type: H-set1 to 5, CQI1 to 30.
Modulation Type: QPSK, 16QAM.
QPSK and 16QAM: Varied according to the combination of FRC type and UE Category.
Refer to Ta ble 5-2 9.
Meas. Type † Sets measurement period mode and measurement period number.
Measurement period mode is selected from TTI and CQI.
Allowable range of measurement period number is from 1 to 100000 in 1 step.
TTI
Measurement period is defined by the TTI period shown by the measurement period number.
The parenthesized value is calculated by "measurement value * 2ms".
CQI
Measurement period is defined by how long it takes before attaining the CQI reporting value
shown by the measurement period number.
H-ARQ retrans Sets the time of retransmission of H-ARQ select 1 or 4.
UE Category Sets UE Category.
Allowable range is from 1 to 6, 11, and 12.
The setting range of Modulation Type of FRC Type and Ec/Ior are varied by the combination of
FRC Type and UE Category.
Refer to Ta ble 5-2 9.
Table 5-32 [Manual Test] Connection Screen Input Field
Input Field Description