BS Level † Sets a RF output level for the tester. The allowable range is from -115.0 to -18.0dBm in 0.1dB
steps.
PWR CNTL Sets control for power control level of the mobile phone to "UP", "DOWN", "CNT UP", or "CNT
DWN" and also a control value.
HOLD: Sets HOLD.
Up: Increases the power control value in the set control value.
Down: Decreases the power control value in the set control value.
CNT UP: Increases the control value continuously.
CNT DWN: decreases the power control value continuously.
The setting is done with the following operations.
When this field is changed from "HOLD" to "UP" or "DOWN", this setting is automatically
returned to "HOLD" after completion of this setting.
When this field is set to "CNT UP" or "CNT DWN", the setting is remained after release the
call.
When this field is changed from "CNT UP" to "CNT DWN", this setting is remained as "CNT
UP" after completion of this setting. (vice versa)
Set TPC Algorithm at "TPC Algorithm" field on the [Configuration: Test Condition] screen.
The allowable range for the control value is from 1 dB to 99 dB in 1 dB step.
Averaging Averages respective measurement values of Frequency Error and EVM.
Set the averaging number from 2 to 99 or "Off".
When this function is not required, select "Off".
In connection status, this setting is available only for BS Call RMC.
BER Frames Sets the number of test frames for BER measurement. The allowable range is from 1 to 4100 in 1
step. (1 Frame = 1 TTI)
Also, a bit number determined from the BER Frames is displayed between parentheses. The bit
number is calculated by multiplying the BER Frame by 244. When the BER Frames is entered, the
bit number is displayed in conjunction with the value of BER Frames
In connection status, this setting is available only for BS Call RMC.
OCNS Sets the OCNS. Select "On" or "Off".
On: Outputs OCNS.
Off: Does not output OCNS.
This is displayed and settable only in the Connection (RMC) state. When the status changes
from BS Call to Connection (RMC), OCNS becomes "Off". Also, this will be reset to "Off" in the
stand-by status even this is set to "On" in the Connection (RMC) status.
CPICH RSCP Sets whether or not to display the value of CPICH RSCP reported from the mobile phone.
Select "On" or "Off".
On: Displays the value of CPICH RSCP and also activates the Measurement CPICH RSCP softkey
in the connection status.
Off: Do not display the value of CPICH RSCP.
If this is set to "On" in the Stand-by status, the value of CPICH RSCP is reported once after MS
or BS call is started and displayed on screen.