User manual
Screen Reference 5
N9360A Multi UE Tester W-CDMA User Manual 5-31
RF Test in Automatic Test
Table 5-4 describes the measurement items of RF test in the
Automatic Test. The limits for measurement items can be set on
the Configuration: Test Condition screen.
Refer to “Configuration: Test Condition Screen" on page 5-156.
Table 5-4 Automatic Test Measurement Items
Measurement Item Description
Open Loop TX Power Measures the power of PRACH.
Refer to “Open Loop Power" on page E-2 for detail.
ILP (Down Min) Measures the minimum variation of the MS output
power level during 10 times of 1 dB down continuously.
Refer to “Inner Loop Power" on page E-9 for detail.
ILP (Down Max) Measures the maximum variation of the MS output
power level during 10 times of 1 dB down continuously.
Refer to “Inner Loop Power" on page E-9 for detail.
ILP (Up Min) Measures the minimum variation of the MS output
power level during 10 times of 1 dB up continuously.
Refer to “Inner Loop Power" on page E-9 for detail.
ILP (Up Max) Measures the maximum variation of the MS output
power level during 10 times of 1 dB up continuously.
Refer to “Inner Loop Power" on page E-9 for detail.
ILP (10 slots Down) Measures the output power level difference between
before 10 dB down and after 10 dB down. Refer to
“Inner Loop Power" on page E-9 for detail.
ILP (10 slots Up) Measures the output power level difference between
before 10 dB up and after 10 dB up. Refer to “Inner Loop
Power" on page E-9 for detail.
MAX TX Power Measures the maximum TX power.
Refer to “Maximum Output Power" on page E-4 for
detail.
Frequency Error Measures the frequency error.
Refer to “Frequency Error" on page E-7 for detail.
EVM Measures the EVM (Error Vector Magnitude).
Refer to “Error Vector Magnitude" on page E-5 for
detail.
Sensitivity/BER Measures the BER (Bit Error Rate) in a weak electric
field. Refer to “Reference Sensitivity" on page E-13 for
detail.
MIN TX Power Measure the minimum TX power. Refer to “Minimum
TX Power" on page E-17 for detail.