User manual
Screen Reference 5
N9360A Multi UE Tester W-CDMA User Manual 5-9
Figure 5-6 Part of [Configuration] screen (without Option E01)
Screen Field
The following input fields need to be set in order to start a new
test. The setting required is described in Table 5-1.
Table 5-1 [Configuration] Screen Input Field
Input Field Description
Reference Selects a reference signal from "Internal" and "External".
• Internal: The internal 10 MHz reference signal is used.
• External: The external 10 MHz reference signal is used. The external 10 MHz reference signal
must be supplied to the "10 MHz Reference IN" connector on the rear panel.
Serial Port Sets the communication condition of the serial port (RS-232C) to use remote control. The option E01 is
required to use this function.
• Baud Rate: Sets communication speed to 9600, 19200, 38400, 57600 or 115200 bits per second.
• Data Length: Sets the data length to 7 or 8 bits.
• Stop Bits: Sets the stop bit length to 1, 1.5 or 2 bits.
• Parity: Sets the parity check mode to "None", "Odd" or "Even".
• Xcontrol: Sets the flow control to "None" or "Xon/Xoff".
GPIB † Sets the communication condition of the GP-IB port to use remote control. The option E00 or option
E02 is required to use this function.
• Addr: Sets the GP-IB address. The allowable range is from 1 to 15.
• EOI: Sets EOI control. Select "On" or "Off".
• On: Enables EOI control. (Terminator setting is ignored.)
• Off: Disables EOI control. (Terminator setting is valid.)
Reboot the Tester after changing the GP-IB setting.
Ter mina tor
• Sets the terminator of the output text data to "CR", "LF" or "CR+LF". This is the setting for remote
control using the serial interface, Ethernet and GP-IB. (The terminator of the input data is always
"LF".
• The Terminator of input text data for GPIB option E00 : set to CR+LF for GPIB option E02 : set to LF.