Technical data

Command Reference 5
Programmer’s Guide 91
Calculate Subsystem
This subsystem is used to perform post-acquisition data processing. In
effect, the collection of new data triggers the CALCulate subsystem.
In this instrument, the primary functions in this subsystem are markers
and limits.
Limit Line Subsection
Limit lines can be defined for your measurement. You can then have the
instrument compare the data to your defined limits and indicate a
pass/fail condition.
Type of Limit Line
:CALCulate:LLINe[1]:TYPE UPPer|LOWer
:CALCulate:LLINe[1]:TYPE?
Sets a limit line to be either an upper or lower type limit line. An upper
line will be used as the maximum allowable value when comparing with
the data.
Example: :CALCulate:LLINe:TYPE UPPer
Key access: Shift > Limit > Limit 1 > Limit Type
Limit Line State
:CALCulate:LLINe[1]|2:STATe OFF|ON|0|1
:CALCulate:LLINe[1]|2:STATe?
Toggles the limit line function between on and off.
Key access: Shift > Limit > Limit 1> Limit Line On/Off
Fixed/Relative Limit
:CALCulate:LLINe:CMODe FIXed|RELative
:CALCulate:LLINe:CMODe?
Toggles the limit line mode between fixed and relative.
Key access: Shift > Limit > Limits Fixed/Rel